9. Special messages, warnings and error indications
Legend: denotes a beeping accompaniment to the message.
denotes alternately flashing messages
.
DISPLAY EXPLANATION REQUIRED ACTION
LO
PROM check sum failed.
power on tests
HI
RAM write/read test failed. cycle power / replace battery,
%
Unidentified power-up failure. if problem persists,
EEPROM
C
heck
S
um
F
ailed. return to your representative
A/D
converter failure.
system tests once per 2 min.
for service.
Ref
erence voltage for A/D out of spec.
Lo
w
bat
tery voltage. replace battery.
LO
TEMP
Instrument
lo
w
temp
erature
rang
e
has been exceeded. make sure that the unit is at a
HI
TEMP
Instrument
hi
gh
temp
erature
rang
e
has been exceeded.
temperature of -1C to +5C.
Sensor circuit is
open
.
dewpoint displays tests
check sensor. wetup sensor to see if it will
react
Sensor is
sat
urated. dry down the sensor.
Sensor circuit is
shor
ted. check sensor.
Trying to calculate dewpoint for undefined sensor. select sensor and autocal.
Viewing dewpoint while sensor is in desiccant
dry
-
storage.
a reminder that the dewpoint in the desic-
cant is being measured.
Trying to calibrate an undefined sensor.
autocal tests
Prompting to
pul
l
out
the sensor from desiccant to
perform autocal.
see autocal instructions.
Sen
sor reading is ‘
too’ lo
w to be from a saturated
sensor, for autocal.
EEP
ROM write cycle not completed.
miscellaneous
if this persists, return for service.
Attempting to modify a
loc
ked unit. unlock unit, see set-up mode.
Sensor is neither
in
nor
out
of desiccant. pull out or push in the sensor actuator
Turn on message,
Xen
taur (
Greek
Ξ
=X)
Turn off message, good-
bye
.
Vue de la page 18