43 Appendix B: Modbus Registers
Holding Registers
Register
Number
Address
Access
Level
Data
Type
Description
Values
Measurement setting registers
40099
98 (62h)
Admin
32-bit
Critical Alarms
bit 0: Dewpoint
bit 1: Sample Temperature
bit 2: Sample Pressure
bit 3: Enclosure Temperature
bit 4: User Input 1
bit 5: User Input 2
bit 6: User Input 3
bit 7: User input 4
bits 8-31: Unused
40101
100 (64h)
User
16-bit
Dewpoint engineering unit
0=°C
1=°F
2=ppmV
3=lbs H
2
O/mm scf
4=g/m
3
40102
101 (65h)
Read-Only
Float
Dewpoint low limit
Valid values are -100 - 99999.9
40104
103 (67h)
Read-Only
Float
Dewpoint high limit
Valid values are -100 - 99999.9
40106
105 (69h)
Admin
Float
Dewpoint low alarm
Valid values are -100 - 99999.9
40108
107 (6Bh)
Admin
Float
Dewpoint high alarm
Valid values are -100 - 99999.9
40110
109 (6Dh)
User
16-bit
Sample temperature engineering unit
0=°C
1=°F
40111
110 (6Eh)
Read-Only
Float
Sample temperature low limit
Valid values are -100 - 100
40113
112 (70h)
Read-Only
Float
Sample temperature high limit
Valid values are -100 - 100
40115
114 (72h)
Admin
Float
Sample temperature low alarm
Valid values are -100 - 100
40117
116 (74h)
Admin
Float
Sample temperature high alarm
Valid values are -100 - 100
40119
118 (76h)
User
16-bit
Sample pressure engineering unit
0=psiA
1=mbar
40120
119 (77h)
Read-Only
Float
Sample pressure low limit
Valid values are 0 - 100
Vue de la page 42